Abstract
We have developed a novel MRI marker for prostate brachytherapy. The purpose of this study was to evaluate the changes in anisotropy when cobalt chloride complex contrast agent encapsulated contrast agent markers (C4-ECAM) were placed adjacent to an iodine-125 (I-125) titanium seed, and to verify that the C4-ECAMs were visible on magnetic resonance imaging (MRI) after radiation exposure. Two C4-ECAMs were verified to be MRI visible in a phantom before radiation exposure. The C4-ECAMs were then attached to each end of a 12.7-U (10-mCi) I-125 titanium seed in a polymer tube. Anisotropy was measured and analyzed with the seed alone and with attached C4-ECAMs by suspending thermoluminescent dosimeters in a water phantom in 2 circles surrounding the radioactive source with radius of 1 or 2 cm. A T1-weighted MRI evaluation of C4-ECAMs was then performed after exposure to the amount of radiation typically delivered during 1 month of prostate brachytherapy. Measured values of the anisotropy function F(r, θ) for the I-125 seed with and without the C4-ECAMs were mutually statistically indistinguishable (standard error of the mean < 4.2%) and agreed well with published TG-43 values for the bare seed. As expected, the anisotropy function φan(r) for the 2 datasets (with and without C4-ECAMs) derived from the measured F(r, θ) did not exhibit statistically measurable difference. Both datasets showed agreement with the published TG-43 φan(r) for the bare seed. The C4-ECAMs were well visualized by MRI after 1 month of radiation exposure. There were no changes in anisotropy when the C4-ECAMs were placed next to an I-125 radioactive seed, and the C4-ECAMs were visualized after radiation exposure. © 2011 American Association of Medical Dosimetrists.
Recommended Citation
Frank, Steven J., Ramesh C. Tailor, Rajat J. Kudchadker, Karen S. Martirosyan, R. Jason Stafford, Andrew M. Elliott, David A. Swanson et al. "Anisotropy characterization of I-125 seed with attached encapsulated cobalt chloride complex contrast agent markers for MRI-based prostate brachytherapy." Medical Dosimetry 36, no. 2 (2011): 200-205. https://doi.org/10.1016/j.meddos.2010.03.004
First Page
200
Last Page
205
Publication Title
Medical Dosimetry
DOI
10.1016/j.meddos.2010.03.004

Comments
Copyright © 2011 American Association of Medical Dosimetrists